Measuring Set: multifunctional instrument for microwave measurements based on sampling digitizer

The basic functions:

  •  Sampling oscilloscope with four independent sweep;
  •  Time-domain reflectometer with 1 cm distance resolution;
  •  Spectrum analyzer based FFT;
  •  Network analyzer( measurements S- parameters, VSWR, characteristic impedancetance);
  •  Time domain dielectric spectroscope;
AppearanceThe block diagram
Appearance The block diagram

Specifications:

 Frequency bandwidth:0 -12, 18, 37 GHz
 Number of channels:2
 Input impedance: 50 Ohm
 Step generator inside
 External clock: up to 100 MHz
 Wide choice additional delivered UWB generator
 Power consumption: 12 V 1,5 A
 Weight: 2 kg
 Dimensions: 210x230x60 mm
 PC interface: LPT, USB 2.0
Measurements parameters:
 Voltage:1 V
 Time interval:from 10 ps to 50 mks
 VSWR:îò 1,05 äî 10
 Attenuation:from 0 to 60 dB
 Reflection coefficient:from 0 to 1
 Wave resistance:from 5 to 200 Ohm
 Distance:from 1 mm to 150 m
Registrator parameters:
 Vertical deflection coefficient: 2 - 200 mV/div
 Sweep deflection coefficient: from 10 ps/div to 5 mks/div
 Voltage measurement accuracy: 1 %
 Time interval measurement accuracy: 0,1 % + 5 ps
 Number of sweep points: 1-2048
 

Options additional delivered UWB generator:

 
Model A(V) (ns) ÀV A Waveform Spectrum
PF-15003-5

PF-1–PF-4
PF-2500,2-0,5

PF-3300,03

PF-460,025

PG-1300,050,15A0,05A PG-1–PG-4
PG-2100,0350,05A0,03A
PG-350,020,05A0,03A
PG-40,20,020,03A0,003A
PF-1.13008

PF-1.1–PF-5.1
PF-2.12004

PF-3.1401

PF-4.1250,5

PF-5.1100,2

  • Generators PG-1 - PG-4 with the output signals in the form of step like function are used in TDR systems, for measurement of a characteristic impedance, a dielectric permittivity.
  • Generators PF1.1-PF5.1 with the output signals in a form of sine wave single period (monocycle) have got a spectral density concentrated near by some central frequency and are used in video pulse location systems as well as in UWB communications when the system bandwidth is limited
  • Generators PF-1-PF-4 with the output signals in a form of Gaussian function with additional small negative wave (monopulse) have the most uniform frequency spectrum and are intended for use as measuring for goals of objects frequency characteristics determination in time domain. They are also used in systems for measuring of broadband antennas (high to low frequency ratio more than 10).
Previous Top Next